| Dr Bernard Reimers is currently Head of the Cathlab in the Department of Cardiology, Mirano General Hospital,Venice, Italy. Prior to this, he was an Assistant in Interventional Cardiology at the Centro Cuore in Milan. Professor Reimers trained, and specialised in Cardiology, at the University of Padova, Italy, and has also worked in both the UK and the Netherlands. |
Use of cerebral protection during carotid artery stenting
Bernhard Reimers
Cardiology Department, Mirano Hospital, Mirano, Italy
Address for correspondence:
Dr Bernhard Reimers
Cardiology Department
Mirano Hospital
Via Mariutto 13, 30035 Mirano, Italy
Tel: +39-(041)-579-4261
Fax: +39-(041)-579-4309
Email: ulss13mir.card@tin.it
Abstract
Carotid artery stenting (CAS) is emerging as a less invasive alternative to carotid endarterectomy, although embolic neurological events occur invariably during CAS. These are often attributable to fragments of atheromatous plaque dislodged during the stenting procedure. Use of a protection device can minimise this risk, and three approaches have been developed: distal occlusive balloons, distal filters and proximal protection by occlusion of common and external carotid arteries. Distal protection devices must cross the lesion for correct positioning and therefore carry a risk of causing embolisation in the absence of cerebral protection. In contrast, proximal systems provide cerebral protection before any device passes the lesion and therefore should enable complete prevention of embolisation. Despite the lack of randomised trials evaluating these devices, available data indicate superiority of protected over unprotected CAS. The capture of debris and prevention of emboli, few device-related complications and technological improvements in newer devices support the use of cerebral protection for CAS.
Introduction
CAS is emerging as an alternative therapy to surgical
carotid endarterectomy (CEA) for the treatment of
extracranial carotid stenosis.1–3 The common goal of
both procedures is the prevention of stroke and the
efficacy of the procedure depends on the periprocedural
complication rates. The endovascular stent procedure
offers a less invasive approach to achieve this goal by
avoiding some of the perioperative complications of
surgical treatment. The randomised trial of carotid
angioplasty (the Carotid and Vertebral Artery
Transluminal Angioplasty Study – CAVATAS) showed
that despite the use of sub-optimal interventional techniques
the early and 3-year outcomes of endovascular treatment or carotid surgery were equivalent.2
Despite the routine application of stents, advanced
stenting techniques, and combined antiplatelet therapy
with aspirin plus clopidogrel or ticlopidine, embolic
neurological events occur invariably during CAS.
Obstructive carotid artery lesions are known to contain
friable, ulcerated and thrombotic material4 that can
embolise during the intervention as shown in
transcranial Doppler, ex vivo5 and in vivo6 studies.
It has also been shown that microembolisation occurs
considerably more frequently during CAS than CEA.7
In order to minimise the risk of embolic neurological
events, a number of protection strategies have been
introduced into the carotid stenting procedure.
A reduction of the Doppler-defined embolic load by
means of a protection device has been demonstrated.8
Preliminary results indicate that the refinement of
stenting techniques, the increasing experience of the
physician, and the routine use of cerebral protection
produce results with CAS similar to the best surgical
series (Figure 1).3,9,10
The first system, a balloon for distal occlusion, was
developed and used by Theron in 1990.11 In current
practice, three different approaches are used for cerebral
protection (Figure 2): two systems of distal protection
(distal occlusive balloons and filters), and proximal
protection using the occlusion of the common and
external carotid arteries. The histopathological analysis
of the debris collected using the various systems of
protection has demonstrated that this debris comprises
fragments of atheromatous plaque dislodged during
carotid stenting.6 In this article, the three
different approaches to cerebral protection are
briefly described.
Distal occlusive balloons
Distal occlusive balloons constitute the first system of
protection used on a large scale.8 They consist of a
guide of 0.014 inches with a balloon on the distal
portion that may be inflated and deflated through a
very small channel contained in the guide itself
(PercuSurge™/GuardWire™). The lesion is crossed with
the guide thereby positioning the balloon distally to the
stenosis where it is inflated until the blood flow in the
internal carotid artery is blocked. Following this, the
angioplasty and stenting procedure is carried out. On
completion of the procedure, a catheter is advanced up
to the distal balloon and the column of blood contained
in the occluded internal carotid artery is aspirated.
In this way debris dislodged incidentally during the
stent procedure is eliminated. Afterwards the balloon is
deflated and the guide is removed. The advantages of distal occlusive balloons are the small diameter (2.2
French) and the good manoeuvrability and flexibility of
the system. Possible disadvantages are that the occlusion
is not tolerated by 6–10% of patients8 and it is not
possible to image the vessel with contrast medium during
the inflation.
Distal filter system
The protection filters consist of a metallic structure
(or skeleton) coated by a membrane of polyethylene
or a net of Nitinol™ wires that contain holes
80–200 µm in diameter.12 The filters are usually
positioned at the distal portion of a 0.014 inch guide.
During the procedure, the filters are folded into a
delivery catheter with which they are advanced distally
to the stenosis. After the lesion is crossed, the filter is opened by removing the delivery sheath. At the end
of the stenting procedure the filter is closed with a
retrieval catheter and is removed from the
carotid artery.
Systems of proximal protection
The distal protection devices – occlusive balloons or
filters – have the disadvantage that they must cross the
lesion before they are inflated or opened.13 This passage
(or crossing) carries the risk of embolisation during this
‘unprotected’ step of the procedure. The proximal
protection systems, in contrast, provide cerebral protection before the passage of any type of device
through the stenoses.
These systems consist of a long introducer sheath with
a balloon that is inflated in the common carotid artery.
A second balloon, inflated in the external carotid artery,
assures the total blockade of the antegrade blood flow
in the internal carotid artery. The proximal protection
systems use the cerebral vascular connections of the
Willis circuit. After the occlusion of the common and
external carotid artery, the collateral flow through the
Willis circuit will create so-called ‘back-pressure’, which will prevent antegrade flow in the internal carotid
artery. After stent positioning, and before the deflation
of the balloons in the common and external carotid
artery, the blood present in the internal carotid artery –
possibly containing dislodged debris – is aspirated
and removed.
The advantage of the proximal protection system is
the fact that the entire procedure is carried out under
protection and, if it is correctly applied, it should
completely avoid any type of embolisation.
The disadvantages of the proximal protection system
are that it is not tolerated by all patients and that the
two systems actually available (Parodi™, Mo.Ma™)
require 10 French introducer sheaths.
Clinical results
In the recent Stenting and Angioplasty with Protection
in Patients at High Risk for Endarterectomy trial
(SAPPHIRE), 156 patients were randomised to carotid
stenting with cerebral protection and 151 patients to
endarterectomy.14 The 30-day incidence of stroke/death
in the group of patients treated by CAS was 4.5% and
in the group treated by CEA was 6.6% (p=not
significant). The SAPPHIRE study clearly shows that
both procedures are at least equivalent in high-risk
patients. If CAS performs at least as well as CEA in
difficult patients, why should it work less well in
patients at ‘normal’ risk? Kastrup et al. reviewed
data from published single-centre reports of 2537
unprotected and 896 protected CAS procedures.9
The 30-day death/stroke rate was 1.8% for patients
treated with protection compared with 5.5% in patients
treated without protection. However, no randomised
trials are available to show the superiority of protected
over unprotected CAS and it is difficult to imagine that
such a randomised controlled study of protected versus unprotected CAS will ever be carried out. Primarily, such
a study would require too many patients to show a
difference between groups, considering the low reported
complication rates of the procedure. However, the
use of these protection devices has some advantages:
firstly, a demonstrated capacity to capture debris which
otherwise would have embolised; secondly, the
reduction of emboli as shown by transcranial Doppler
studies;8 thirdly, an increase in experience using these
devices with a low incidence of device-related
complications;3 and finally, the availability of
second-generation devices featuring much higher
flexibility, lower crossing profile, and easier handling.
Conclusions
Carotid artery stenting is emerging as an alternative,
less invasive therapy for the treatment of carotid
stenoses. Cerebral protection devices have been used
successfully in large series of CAS and low
periprocedural complication rates have been obtained.
Based on the documented feasibility and safety of
protection-device handling3 and the previously shown
capacity of protection devices to reduce embolisation
of debris into the cerebral circulation,6 it appears
prudent to consider cerebral protection mandatory
for any carotid artery stent procedure.
With respect to the controversy surrounding
endovascular versus surgical therapy, the results of the
first randomised studies and data from the registers of
CAS with cerebral protection indicate outcomes similar
to those obtained from the best surveys of
surgical endarterectomy.
Key Learning
• Embolic neurological events occur invariably during carotid artery stenting (CAS) despite
advances
in technique, technology and anti-thrombotic regimens
• The risk of embolic neurological events can be minimised using cerebral protection
devices:
– distal occlusive balloons
– distal filter systems
– proximal protection systems (occlusion of the common and external carotid arteries)
• Distal devices have to cross atheromatous plaque before deployment, increasing the
risk of
embolisation without cerebral protection in place
• A proximal protection system can provide cerebral protection from any type of
embolisation
throughout the procedure
• Available data support the use of cerebral protection devices during CAS |
References
1. Roubin SG, New G, Iyer SS, et al. Immediate and late clinical
outcomes of carotid artery stenting in patients with symptomatic and
asymptomatic carotid artery stenosis. A 5-year prospective analysis.
Circulation 2001;103:532–7.
2. Endovascular versus surgical treatment in patients with carotid
stenosis in the carotid and vertebral artery transluminal angioplasty
study (CAVATAS): a randomised trial. Lancet 2001;357:1729–37.
3. Reimers B, Schlüter M, Castriota F, et al. Routine use of cerebral
protection during carotid artery stenting: results of a multicenter
registry of 753 patients. Am J Med 2004;116:217–22.
4. Imparato AM, Riles TS, Gorstein F. The carotid bifurcation plaque:
pathologic findings associated with cerebral ischemia. Stroke
1979;10:238–45.
5. Ohki T, Roubin GS,Veith FJ, et al. Efficacy of a filter device in the
prevention of embolic events during carotid angioplasty and stenting:
an ex vivo analysis. J Vasc Surg 1999;30:1034–44.
6. Angelini A, Reimers B, Dalla Barbera M, et al. Cerebral protection during
carotid artery stenting: collection and histopathologic analysis of
embolized debris. Stroke 2002;33:456–61.
7. Crawley F, Clifton A, Buckenham T, et al. Comparison of hemodynamic
cerebral ischemia and microembolic signals detected during carotid
endarterectomy and carotid angioplasty. Stroke 1997;28:2460–4.
8. Al-Mubarak N, Roubin GS,Vitek JJ, et al. Effect of the distal-balloon
protection system on microembolization during carotid stenting.
Circulation 2001;104:1999–2002.
9. Kastrup A, Groschel K, Krapf H, et al. Early outcome of carotid
angioplasty and stenting with and without cerebral protection devices:
a systematic review of the literature. Stroke 2003;34:1936–41.
10. Brooks WH, McClure RR, Jones MR, et al. Carotid angioplasty and
stenting versus carotid endarterectomy: randomized trial in a community
hospital. J Am Coll Cardiol 2001;38:1589–95.
11. Theron J, Courtheoux P, Alachkar F, et al. New triple coaxial catheter
system for carotid angioplasty with cerebral protection.
Am J Neuroradiol 1990;11:869–74.
12. Reimers B, Corvaja N, Moshiri S, et al. Cerebral protection with filter
devices during carotid artery stenting. Circulation 2001;104:12–5.
13. Adami CA, Scuro A, Spinamano L, et al. Use of the Parodi
anti-embolism system in carotid stenting: Italian trial results. J Endovasc
Ther 2002;9:147–54.
14.Yadav J. Oral presentation at the 75th Annual Scientific Sessions of
the American Heart Association, Chicago, November 2002.
|