Carotid artery stenting without cerebral protection
Arnd Doerfler, Tobias Engelhorn and Michael Forsting
Department of Neuroradiology, Diagnostic and
Interventional Radiology, University of Essen Medical
School, Essen, Germany
Address for correspondence:
Dr Arnd Doerfler, Department of Neuroradiology
Diagnostic and Interventional Radiology
University of Essen Medical School
Hufelandstrasse 55, D-45122 Essen, Germany
Tel: +49-(201)-723-1542
Fax: +49-(201)-723-5959
Email: arnd.doerfler@uni-essen.de
Abstract
Carotid angioplasty and stenting is increasingly used as an attractive alternative to carotid endarterectomy (CEA) for treatment of symptomatic and asymptomatic carotid artery disease. Results from observational, non-randomised studies describe excellent technical success and periprocedural complication rates similar to those of CEA. Technical success, patency and improved complication rates are mainly the result of substantial technical improvements, an improved associated learning curve, and the improved periprocedural anti-coagulation regimens.
Currently, controversies exist as to whether carotid angioplasty with stenting should be carried out with or without obligatory use of cerebral protection. In experienced hands, the use of protection devices can decrease complication rates. However, protection systems increase catheter time and technical complexity. Reviewing the literature, the combined stroke and death rate in protected carotid angioplasty with stenting is 2.0% compared with 3.2% in unprotected carotid angioplasty with stenting, which does not strongly support the use of protection devices. Additionally, scientific evidence for the mandatory use of protection devices is lacking. Until the results of trials comparing CEA and carotid angioplasty with stenting are available, carotid artery stenting is indicated only in selected patients in an interdisciplinary approach and at high-volume centres, to obtain maximal patient benefit.
Percutaneous carotid angioplasty with stenting
During the early 1990s, the North American
Symptomatic Carotid Endarterectomy Trial (NASCET) in
symptomatic patients and the Asymptomatic Carotid
Atherosclerosis Study (ACAS) trial in asymptomatic
patients established carotid endarterectomy (CEA)
as the standard of care in patients with
high-grade carotid artery stenosis [1,2].
Percutaneous carotid angioplasty with stenting has
the advantage of being less invasive (and thus less
traumatic) but usable in patients considered to be at
high surgical risk, and may therefore offer an attractive
therapeutic alternative to CEA in selected patients.
Observational evidence is increasing rapidly to support
the widespread use of carotid angioplasty with stenting.
Results from non-randomised studies describe good
technical success, and periprocedural complication
rates and long-term benefits similar to those of CEA [3-5]. An updated worldwide survey including more than
12,000 procedures carried out at 36 medical centres
reported a technical success rate of carotid angioplasty
with stenting close to 99%, a complication rate for
stroke and death of <5%, and a restenosis rate of 2.4%
at 3 years, respectively.6 However, the quality of data
collection at each of these centres is difficult to
ascertain and this type of survey information is subject
to the bias of those reporting the information, almost
certainly leading to an under-reporting of complications.
Sufficient data from randomised clinical trials
comparing carotid angioplasty with stenting and CEA
are still not available. The Carotid and Vertebral Artery
Transluminal Angioplasty Study (CAVATAS) was the first
prospective, multicentre, randomised trial comparing
carotid endovascular intervention with CEA. The
interventions were carried out by operators in the
learning curve, without protection devices, with
contemporary appreciation for the use of aggressive
anti-platelet therapy, and using state-of-the-art
technical equipment. Periprocedural complication rates
were equivalent for both procedures but were significantly higher than those seen in the single-centre
studies carried out during the period 1990-1999 and
those reported in NASCET and ACAS. In CAVATAS, most
of the carotid angioplasty with stenting patients were
treated by angioplasty alone, and stenting was
conducted in only 26% of endovascular patients [7].
The marked improvement of carotid angioplasty with
stenting in the reduction in complications and improved
clinical outcome is mainly the result of
three factors:
(i) Recent substantial technical improvements - initially,
the stent devices used were mainly adapted from
coronary or peripheral vascular applications. Today,
dedicated carotid stenting equipment with low-profile
stent delivery systems, with a variety of different
self-expandable stent designs, better access sheaths
and specially designed wires and low profile balloons,
are available.
(ii) Increased operator experience - as with any
interventional procedure, outcomes are influenced by
operator expertise and experience. The learning curve
for active interventionalists beginning carotid
angioplasty with stenting is very steep.
(iii) The improved periprocedural anti-coagulation
regimens - using combined platelet inhibitors, such
as acetylsalicylic acid and clopidogrel, low-molecular
weight heparin before, and full heparinisation during
the procedure.
Evaluation of carotid cerebral protection devices
Currently, the dominant point of discussion centres on
carotid cerebral protection devices, with controversy
existing as to whether carotid angioplasty with
stenting should be carried out with or without
cerebral protection.
At first glance, it seems reasonable to apply protection
systems to catch particles, by means of occlusive
balloon systems or filtration baskets in the internal carotid artery, before they reach the cerebral circulation,
thereby minimising the risk of associated embolic
neurological complications. The beneficial use of such
devices seems to be supported by an increasing
number of publications reporting reduced neurological
complications. Theron et al. were the first to promote
use of a distal protection device made of a triple
co-axial catheter, which enabled carotid angioplasty
with temporary internal carotid artery (ICA) occlusion,
aspiration of debris, and flushing into the external
carotid artery [8]. Henry et al. further improved this
technique with a commercially available low-profile
occlusion balloon system [9]. Parodi et al. focused on
proximal embolic protection by temporarily reversing
flow in the ICA [10], a technique that needs an [11] French
femoral access. Recently, Cremonesi et al. reported their
single-centre experience with protected carotid angioplasty
with stenting in 442 patients with an overall
complication rate of 3.4%, and a 30-day ipsilateral
stroke/death rate of 1.1%, concluding that protective
devices are effective in preventing distal embolisation [11].
However, the published data come from uncontrolled
trials with variable patient selection and are collected
retrospectively. Alternatively, the collection of data is
not otherwise unbiased, for example, in industrysponsored
trials or where the investigators are affiliated
to the companies developing or distributing the devices.
Disadvantages of cerebral protection systems
Cerebral protection systems have themselves some inherent disadvantages, namely an increase in both catheter time and technical complexity, both of which might be associated with an increased complication rate. The advantage of filter systems is that they provide continuous perfusion to the brain.
The disadvantages for some systems include a larger diameter than distal balloon occlusions, difficulties in tracking more angulated bifurcations and tortuous vessels, and the theoretical issue of missing very small particles. Balloon devices comprise the second group of embolic protection devices. The principal disadvantage of these is that 5-10% of patients do not tolerate carotid occlusion [9]. Rarely, significant dissection of the internal carotid artery occurs. Another disadvantage is that angiography during the procedure is not possible.
The primary passage of the stenosis must always be an unprotected manoeuvre. Passing more instrumentation across friable plaque can itself lead to complications [11], such as cerebral embolism, vasospasm or, rarely, dissection. Primary stenting, i.e. stent implantation before the first balloon inflation, may also help to prevent distal embolisation by fixation of the debris to the vessel wall. However, clinically relevant embolisation may still occur. As a result of the larger diameter of the protection devices, pre-dilatation of the stenosis is often necessary before the protection device may be placed. Transcranial Doppler imaging studies report pre-dilatation as the most dangerous step of carotid angioplasty with stenting [18]. In the study of Cremonesi et al [11]. pre-dilatation was necessary in 37% of patients compared with only 2% of patients for unprotected carotid angioplasty with stenting [11,19]. After stent placement and post-dilatation, removal of protection devices may cause additional microembolisation [21]. Thus, from a procedural point of view, protection devices may reduce, but do certainly not eliminate, plaque embolisation.
A requirement for data from comparative trials
In addition to the factors discussed above, quality of evidence for the use of protection devices is still lacking [20]. As a result of the low complication rate of carotid angioplasty, even without cerebral embolic protection devices, large randomised trials are necessary to prove the clinical efficacy of these devices. Thus, all efforts should concentrate on finishing the ongoing trials comparing CEA and carotid angioplasty with stenting†, such as SPACE (Germany), CAVATAS (United Kingdom), EVA-3S (France) and CREST (United States). It is anticipated that these trials will publish definitive results within the next 1-3 years and help guide the referral of patients for carotid angioplasty with stenting and CEA in the future. Until these results are available, no scientific evidence is available to support the use of carotid angioplasty with stenting over CEA in appropriate patients or the use of carotid angioplasty with stenting in asymptomatic patients.
Indications for carotid angioplasty with stenting
In our opinion, the only generally accepted indication for carotid angioplasty with stenting outside of clinical trials is patients with medical contraindications to CEA, post-radiation carotid stenosis, restenosis of a previous CEA, high cervical anatomically inaccessible lesions, or contralateral carotid occlusions.
Importantly, carotid angioplasty should be carried out as an interdisciplinary approach in centres with a sufficient case-load where an experienced team of interventional neuroradiologists and neurologists closely collaborate to obtain maximal patient benefit.
- Carotid angioplasty with stenting is increasingly used as a less invasive alternative to carotid endarterectomy in patients with carotid artery disease
- Improvements in technology, anti-coagulation regimens and operator experience have reduced complication rates and improved the clinical outcome from carotid angioplasty with stenting
- The use of cerebral protection devices during carotid angioplasty with stenting is controversial and scientific evidence for the obligatory use is still lacking
- Use of cerebral protection devices has disadvantages including: – increased catheter time
- Carotid angioplasty with stenting can be successfully carried out without cerebral protection in routine clinical practice
- Ongoing clinical trials will provide better evidence on the clinical efficacy of cerebral protection devices
– increased technical complexity
– in some cases pre-dilatation of stenosis with the risk of plaque embolisation
References
- Beneficial effect of carotid endarterectomy in symptomatic patients with high-grade carotid stenosis. North American Symptomatic Carotid Endarterectomy Trial Collaborators. N Engl J Med 1991;325:445-53.
- Asymptomatic Carotid Atherosclerosis Study Executive Committee: endarterectomy for asymptomatic carotid artery stenosis. JAMA 1995;273:1421-28.
- Yadav JS, Roubin GS, Iyer S, et al. Elective stenting of the extracranial carotid arteries. Circulation 1997;95:376-81.
- Roubin GS, New G, Iyer SS, et al. Immediate and late clinical outcomes of carotid artery stenting in patients with symptomatic and asymptomatic carotid artery stenosis: a 5-year prospective analysis. Circulation 2001;103:532-7.
- Brooks WH, McClure RR, Jones MR, et al. Carotid angioplasty and stenting versus carotid endarterectomy for treatment of asymptomatic carotid stenosis: a randomized trial in a community hospital. Neurosurgery 2004;54:318-24.
- Wholey MH, Al-Mubarek N. Updated review of the global carotid artery stent registry. Catheter Cardiovasc Interv 2003;60:259-66.
- Brown MM, Rogers J, Bland JM, the CAVATAS Investigators. Endovascular versus surgical treatment in patients with carotid stenosis in the carotid and vertebral artery transluminal angioplasty study (CAVATAS): a randomized trial. Lancet 2001;357:1729-37.
- Theron JG, Payelle GG, Coskun O, et al. Carotid artery stenosis: treatment with protected balloon angioplasty and stent placement. Radiology 1996;201:627-636.
- Henry M, Amor M, Henry I, et al. Carotid stenting with cerebral protection: first clinical experience using the PercuSurge GuardWire system. J Endovasc Surg 1999;6:321-31.
- Parodi JC, La Mura R, Ferreira LM, et al. Initial evaluation of carotid angioplasty and stenting with three different cerebral protection devices. J Vasc Surg 2000;32:1127-36.
- Cremonesi A, Manetti R, Setacci F, et al. Protected carotid stenting: clinical advantages and complications of embolic protection devices in 442 consecutive patients. Stroke 2003;34:1936-41.
- Kastrup A, Groschel K, Krapf H, et al. Early outcome of carotid angioplasty and stenting with and without cerebral protection devices: a systematic review of the literature. Stroke 2003;34:813-9.
- Bonaldi G. Angioplasty and stenting of the cervical carotid bifurcation: report of a 4-year series. Neuroradiology 2002;44:164-74.
- Criado FJ, Lingelbach JM, Ledesma DF, et al. Carotid artery stenting in a vascular surgery practice. J Vasc Surg 2002;35:430-4.
- Hobson RW 2nd, Lal BK, Chakhtoura EY, et al. Carotid artery closure for endarterectomy does not influence results of angioplasty-stenting for restenosis. J Vasc Surg 2002;35:435-8.
- Jaeger H, Mathias K, Drescher R, et al. Clinical results of cerebral protection with a filter device during stent implantation of the carotid artery. Cardiovasc Intervent Radiol 2001;24:249-56.
- Qureshi AI, Suri MF, New G, et al. Multicenter study of the feasibility and safety of using the memotherm carotid arterial stent for extracranial carotid artery stenosis. J Neurosurg 2002;96:830-6.
- Orlandi G, Fanucchi S, Fioretti C, et al. Characteristics of cerebral microembolism during carotid stenting and angioplasty alone. Arch Neurol 2001;58:1410-3.
- Koch C, Kucinski T, Eckert B, et al. Endovascular therapy of high-degree stenoses of the neck vessels-stent-supported percutaneous angioplasty of the carotid artery without cerebral protection. Fortschr Rontgenstr 2002;174:1506-10.
- Eckert B, Zeumer H. Editorial comment: Carotid artery stenting with or without protection devices? Strong opinions, poor evidence! Stroke 2003;34:1941-3.
- Al-Mubarak N, Roubin GS,Vitek JJ, et al. Effect of the distal-balloon protection system on microembolization during carotid stenting. Circulation 2001;104:1999-2002.
December 2004, 1098/OS




Use of cerebral protection during carotid artery stenting
The balance of evidence for carotid artery stenting with cerebral protection